Specifications :
  • Dimensions = 50cm x 23cm x 32cm,
  • Weight = 19kg (42lbs) - without accessories. Control Ranges
  • WORKING PRESSURE =Up to 120cm H2O.
  • MODES: Volume controlled, volume control with sight, pressure controlled, pressure support, synchronized intermittent mandatory ventilation (SIMV),
  • SIMV+ pressure support, continuous positive airway pressure, manual ventilation.
  • Present inspiratory minute volume: .5 - 40 1/min, with sufficient working pressure.
  • FLOW PATTERN: constant, accelerating or decelerating.
  • RESPIRATORY RATE: 5-120 breaths/min.
  • INSPRIATORY TIME: 20-80% of respiratory cycle.
  • PAUSE TIME: 0-30% of respiratory cycle.
  • PEEP: 0-50cm H2O.
  • UPPER LIMIT: 15-120cm H2O
  • TRIG SENSITIVITY: -20-0cm H2O (below PEEP)
  • SIMV RATE: high rate (4-40 SIMV breaths/min), low rate (.4-4 SIMV breaths/min).
  • SIGHT: double tidal volume possible every 100 breaths.
  • COMPRESSED VOLUME: 0.075 1(internal).
  • Monitoring
  • ANALOG PRESENTATION:
  • AIRWAY PRESSURE: -20 - 120cm H2O
  • EXPIRED MINUTE VOL: adults 0-40 1/min, infants 0-4 1/min.
  • DIGITAL PRESENTATIONS:
  • RESPIRATORY RATE: 5-120 breaths/min.
  • O2 CONCENTATION: in % by volume, 20-100%.
  • INSPIRED TIDAL VOLUME: 0-1999 ml.
  • EXPIRED TIDAL VOLUME: 0-1999ml.
  • EXPIRED MINUTE VOLUME: adults 0-40 1/min, infants 0-4 1/min.
  • PEAK PRESSURE: 0-120cm H2O
  • PAUSE PRESSURE: 0-120cm H2O
  • MEAN AIRWAY PRESSURE: 0-120CM H2O
  • Gas & Power Supply
  • Must fulfil specifications for medical gases.
  • GAS CONSUMPTION: approximately equal to preset inspiratory minute volume or equal to flow meter setting.
  • GAS SUPPLY VIA HIGH PRESSURE INLET: from central gas supply, compressor or gas cylinder.
  • ELECTRONIC GAS SUPPLY UNIT, INLET PRESSURE: 2.5-5 bar (250-500 kPa; 35-70 psi).
  • OTHER GAS SUPPLY INLET PRESSURE: 2.5-7 bar (250-700 kPa; 35-100 psi).
  • GAS SUPPLY VIA LOW PRESSURE INLET: vial flow meters.
  • POWER SUPPLY: (100), 110, 120, 220, 240, 50 OR 60 Hz.
  • POWER CONSUMPTION: Approximately 40 W. Alarms & Indications
  • PUSH BUTTON: audible signals, muted for 2 min maximum.
  • ALARMS: gas supply alarm, apnea alarm, expired minute volume, O2 concentration, upper pressure limit, power failure.
  • INDICATORS: set minute volume alarm, set O2 alarm, set O2 alarm, pressure support, CPAP, MAN SIMV high/low rate, expired minute volume infants, patient trig.
  • Special Functions :
  • Inspiratory pause hold, expiratory pause hold gas exchange.